There are several options, such as:
1. Look for a 3rd edition compliant power supply from a different supplier.
2. If only a Means of Operator Protection (MOOP) is required: it is allowed to use a power supply that complies with EN 60950-1:2005. In addition to the barrier requirements for MOOP according EN 60950, the 3rd edition contains additionally applicable requirements like leakage currents etc.
3. If MOPP is required: If the MEE or MES contains a 2Ed approved power supply, then this power supply needs to undergo a delta testing to 3Ed 60601-1.
Do you have a question ask it here?
Leaders in EMC Testing , Environmental Testing, Safety Testing, Radio and Telecoms Testing, CE Marking, Vibration Testing, Climatic Testing, Certification, Training and Consulting.
Leaders in EMC Testing , Environmental Testing, Safety Testing, Radio and Telecoms Testing, CE Marking, Vibration Testing, Climatic Testing, Certification, Training and Consulting.